Read Commands "Cvc 2000 - vacuubrand ME 4C NT VARIO Instructions For Use Manual

Chemistry diaphragm pumps. chemistry vacuum system. chemistry pumping units
Hide thumbs Also See for ME 4C NT VARIO:
Table of Contents

Advertisement

Read commands "CVC 2000"

Command
Operation
IN_PV_1
current pressure
IN_PV_2
current frequency
device set
IN_CFG
preselections
IN_ERR
error code
status of
IN_STAT
process control
page 84 of 129
Response
Description
XXXX mbar/
unit according to preselections
Torr/hPa
XX.X Hz
pump speed
0XXXX
VACUU•LAN
1XXXX
continuous pumping
2XXXX
vacuum control without automatic
3XXXX
vacuum control with automatic
X0XXX
no coolant valve
X1XXX
coolant valve
XX0XX
no venting valve
XX1XX
venting valve
XXX0X
no automatic switch off
XXX1X
automatic switch off
XXXX0
remote operation off
XXXX1
remote operation on
1XXX
fault at pump electronics
X1XX
overpressure
XX1X
maloperation mode pressure transducer
XXX1
last command to interface incorrect
coolant valve closed
0XXX
coolant valve open
1XXX
venting valve closed
X0XX
venting valve open
X1XX
VACUU•LAN: inactive
XX00
VACUU•LAN: pumping down,
XX01
current pressure > selected pressure
XX02
VACUU•LAN: pumping down, time for
automatic switching off is running
XX03
VACUU•LAN: system is switched off
XX10
continuous pumping: not active
XX11
continuous pumping; active
XX20
vacuum control: not active
XX21
vacuum control:
current pressure above set vacuum
XX22
vacuum control: current pressure
equals set vacuum (±1 Torr/mbar)
XX23
vacuum control:
current pressure below set vacuum
XX30
automatic: not active
XX31
automatic: determining boiling point
XX32
automatic: adjusting boiling point
XX33
automatic: system is switched off