Cleaning and decontaminating

IN
sartorius stedim
121°C
2 bar absolute
max.
80%
121°C
2 bar
absolute
max. 20 x
UV disinfection is permitted but may lead to discoloration
NOTICE
of the plastic parts.
Attention: UV disinfection acts only at the surface.
page 21 of 35
Autoclaving
The collecting bottle with bottle head and screw
cap, the quick coupling and the filter are des-
ignated for steam sterilization at 121°C and 2
bar absolute (1 bar overpressure). Time of ex-
posure according to DIN 58946 t
NOTICE
Prior to autoclaving loosen or remove the
bottle head from the bottle.
It is the user´s responsibility to ensure effective
autoclaving.
NOTICE
Autoclaving
Maximum liquid level in collecting bottle:
80%. When working with low boiling liquids or
liquids that tend to foam, the maximum liquid
level may be less.
The number of autoclaving cycles can be
marked on the plastic disc (Memory Disc) of the
filter (max. 20 autoclaving cycles according to
the manufacturer instruction).
= 20 minutes.
e