Beckman Coulter SX4750A Instruction For Use, Installation And Assembly page 23

Swinging-bucket rotor
Table of Contents

Advertisement

Figure 3 Blood Bag Loading Procedures
2
1
1. Blood Bag Label
2. Blood Bag Seam Aligned with
Pivotal Axis of Bucket
3. Ports
3
Make sure the loaded blood bag cups opposite each other on the rotor yoke are approximately
the same weight.
• Soft, pliable balancing pads that will not puncture blood bags (see the
used to maintain weight balance, if necessary.
4
Place loaded cups into rotor buckets.
• To reduce the possibility of bag breakage, align the blood bag seam with the rotor pivot pins
(see
5
Place bucket covers on rotor buckets to prevent blood bags or tubing from contacting the rotor
yoke during centrifugation.
GX-TB-004EE
1
1
4. Blood Bag Seam
5. Blood Bag Cup
6. Sandwich Coiled Tubing between
Figure
3) with the label facing out.
6
1
Transfer Bags and Filled Blood Bag
SX4750A Swinging-Bucket Rotor
Loading the Buckets
3
4
1
5
List) can be
Supply
1-13