Operating The Full Length Side Rails - Ultimate Healthcare Cura II User Manual

Community bed frames
Table of Contents

Advertisement

10. Undo and remove the hand screw at the bottom of each fixing channel
11. The next section is easier if you extend the bed platform slightly, follow the instructions
in section 12 earlier in this manual.
12. Ensure the end of fixing wire is secured into the hole shown on the image below
13. Slide the plastic side rail inserts into the inner part of the sliding bar and secure by
reinserting the screw
14. Slot the rails into the inserts starting with the bottom rail and repeat on the other side.
15. Once all full length side rails have been attached ensure you return the bed frame to its
standard length. This will fix the side rails in place.
16. Test the side rails operate correctly.
17. Ensure that the distance from the top of the uncompressed mattress and the top
of the side rail is at least 220mm.
13.3.1
1. Raise the side rail until each end clips into place.
2. Operate the clip on the side of the sliding bar bracket to lower the side rail. The side rail
may need to be raised slightly to operate the clip.
Issue: 6: 08/11/2018
DRAFT
Hand screw
Fixing wire
Hole to insert fixing wire

Operating the Full Length Side Rails

Page 25 of 40