Appendix C. Wiring Charts; Millivolts Per Volt Output; Zero To Five Volt Dc Output; Four To 20 Milliamp Output - Geokon 3400 Series Instruction Manual

Semiconductor piezometer
Table of Contents

Advertisement

APPENDIX C. WIRING CHARTS

C.1 Millivolts per Volt Output
Geokon Cable
#04-375V9 (Violet)
Red
Red's Black
White
White's Black
Green
Green's Black
Blue
Blue's Black
Shields (5)
Note:
Input voltage for Model # 3400-1, mV/V output is 10V d.c.
(Power -, Signal -, Remote Sense -, are connected internally.)
C.2 Zero to Five Volt DC Output
Geokon Cable
#04-375V9 (Violet)
Red
Red's Black
White
White's Black
Blue
Blue's Black
Shields (5)
Note:
Input voltage for Model # 3400-2, 0–5VDC output is 6.5–35V d.c.
C.3 Four to 20 Milliamp Output
Geokon Cable
#02-250V6 (Blue)
Red
Black
White
Green
Shields (1)
Note:
Input voltage for Model # 3400-3, 4–20mA output is 6.5–35V d.c.
Internal Sensor Wiring
Red
Black
White
Black
Red
Black
N/C
N/C
N/C
Table 6 - mV/V Output Wiring
Internal Sensor Wiring
Red
Black
White
Black
N/C
N/C
N/C
Table 7 - 0-5VDC Output Wiring
Internal Sensor Wiring
Red
Black
N/C
N/C
N/C
Table 8 - 4-20mA Output Wiring
Function/Description
Power +
Power -
Signal +
Signal -
Remote Sense +
Remote Sense -
Thermistor
Thermistor
Ground
Function/Description
Power +
Power -
Signal +
Signal -
Thermistor
Thermistor
Ground
Function/Description
Power +
Power -
Thermistor
Thermistor
Ground
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

3400dp3400h