Safety notes
1.5

Organisational Measures

Page 8
1.5
Organisational Measures
 Before using the device, ensure that a medical product representative has
explained its functions as well as the safety requirements.
 Keep this user guide in an accessible place for reference purposes. Make sure
that it is always complete and legible.
 Observe the operating and maintenance instructions.
 These operating instructions do not override any statutory or local regulations, or
procedures for occupational safety and environmental protection.
1.6

Safety-conscious Operation

 Only operate the device in accordance with the specified technical data.
©
The device is CF classified. It is defibrillation protected only when the
SCHILLER original patient cable is used. However, as a safety precaution and if
possible, remove the electrodes before defibrillation.
 Do not touch the unit during defibrillation.
 To ensure patient safety, none of the electrodes including the neutral electrode,
nor the patient or any person with simultaneous patient contact, must come in
contact with conductive parts, even when these are earthed.
 Immediately report any changes that impair safety (including operating
behaviour) to the responsible person.
 Do not place any liquids on the unit. If liquid is spilled on the device, immediately
disconnect the device from the mains and wipe it. The device must be checked
before reusing.
 Only connect the original SCHILLER patient cable to the patient socket.
 If the patient cable should become defective after defibrillation, an electrode
becomes displaced, or an electrode resistance is too high, a lead-off indication is
displayed in the upper right part of the screen (see
patient cable check, page 25).
 Only use accessories and disposables recommended or supplied by SCHILLER.
The use of third-party accessories (including disposables) may result in injury,
inaccurate information due to electromagnetic interferences and/or damage to
the device.
1.7

Safety facilities

 Operating the device without the correctly rated fuse or with defective cables
constitutes a danger to life! Therefore:
– Do not operate the unit if the earth connection is suspect or if the mains lead,
the power supply unit or the device is damaged or suspected of being damaged.
– Damaged cable connections and connectors must be replaced immediately.
– Electrical safety devices, such as fuses, must not be modified.
– Fuses must only be replaced with the same type and rating as the original.
CARDIOVIT AT-1 G2