Advertisement

3

Operation

3.1

Initial operation

Page 16
3 Operation
3.1
Initial operation
 Electrical shock hazard. Do not operate the unit if the earth connection is suspect
or if the power supply unit/mains lead is damaged or suspected of being
damaged.
3.1.1

Location

• Do not keep or operate the unit in a wet, moist or dusty environment. Avoid expo-
sure to direct sunlight or heat from other sources.
• Do not allow the unit to come into contact with acidic vapours or liquids.
• The CARDIOVIT AT-1 G2 should not be placed in the vicinity of X-ray or diathermy
units, large transformers or electric motors.
3.2

Connections

 All externally connected hardware must be approved by SCHILLER. Connection
of any hardware not approved by SCHILLER is at the owner's risk. Moreover, the
unit's warranty may become invalid.
 Position the device so that the mains connection (1) is easily accessible at all
times.
3.2.1

Back panel

1
(1) Mains connection 100...240 VAC
(2) Potential equalisation stud. The potential equalisation stud is used to equalise
the ground potential of the unit to that of any nearby mains-powered equipment.
Use the hospital or building common ground for all mains-powered units.
(3) USB interface for USB memory stick.
(4) Kensington lock
3
2
CARDIOVIT AT-1 G2
4