Baudrate; Rs 232 C Data Output Via Interface Rs 232 C; General Information; Data Output Via Print-Key - KERN 572 Operating Instructions Manual

Electronic precision and platform balances
Table of Contents

Advertisement

5.3

Baudrate

The Baud rate for the data transfer is adjusted with the MODE-key. The following example
demonstrates how to set the Baud rate 4800.
Select Baud rate
1. Press MODE-key repeatedly until "PRINTER" is displayed.
2. Press YES-key.
3. Press MODE-key repeatedly until the desired Baud rate
appears (for instance 4800 Baud).
4. Press YES-key to select 4800 Baud. The tick-mark (X)
confirms the new setting.
5. Press MODE-key repeatedly until the balance displays in
grams again, or press tare key.
5.4

RS 232 C Data output via interface RS 232 C

5.4.1 General information

As a condition for the data transfer between the balance and a peripheral device (for
instance printer, PC ...) both devise have to be set on the same interface parameter (for
instance baud rate, parity ...).
There are 4 methods for the data output via RS 232 C
1. Data output via PRINT-key (see 5.4.2).
2. Data output, after having loaded the balance, by setting AUTOPRINT (see 5.4.3).
3. Continuous data output by setting AUTOPRINT PC (see 5.4.4).
4. Data output by transfer of remote controls (see 5.4.5)

5.4.2 Data output via PRINT-Key

The printing process can be released by the PRINT-key. In this case the settings
AUTOPRINT and AUTPRINT PC should be deselected.

5.4.3 AUTOPRINT

The setting AUTOPRINT is in the PRINTER-routine, and there it can be selected or
deselected. When AUTOPRINT is active the actual weighing value will be sent via the RS
232 interface when the balance has been unloaded and then loaded after having achieved
the stability.
36
4800 Baud X
572/KB/DS-BA-defsi-0246
Display
PRINTER?
2400 Baud
4800 Baud
0,0 g