Mechanical Specifications; Electrical Specifications - MSA FL500 Operating Manual

Uv/ir flame detector
Hide thumbs Also See for FL500:
Table of Contents

Advertisement

8.2

Mechanical Specifications

Enclosure material
Color
Finish
Height
Width
Depth
Weight
Cable Entry
Earth Terminals
8.3

Electrical Specifications

Nominal supply voltage
Supply voltage range
Maximum supply current
Typical current
Maximum output signal load 600 ohms
Output signal range
Fault signal
COPM fault signal
Ready signal
IR only signal
UV only signal
Alarm Low signal
Alarm High signal
Relay contact ratings
RS-485 output
HART
HART impedance
316 Stainless Steel
Red
TGIC polyester powder coat
6.4 in. (16.2 cm)
5.0 in. (12.7 cm)
4.3 in. (11.0 cm)
9.4 lb (4.3 kg)
2 x ¾ in. NPT (M25 adapters included for some config-
urations)
8-32 x 3/8 Philips pan-head screw, stainless steel,
zinc-plated, green dye; external stainless steel tooth
lock washer included. Applicable for connection of
2
4 mm
or smaller.
24 Vdc
20 to 36 Vdc (measured at device)
200 mA during COPM only
80 to 150 mA
0 to 20 mA *
0 to 0.2 mA *
2.0 ±0.2 mA *
4.0 ±0.2 mA
8.0 ±0.2 mA
12.0 ±0.2 mA
16.0 ±0.2 mA
20.0 ±0.2 mA
North American Approved Applications:
SPDT, 5A @ 250 Vac, or 5A @ 30 Vdc resistive max.
European Union (EU) Approved Applications:
SPDT, 5A 30V RMS/42.4V Peak, or 5A @ 30 Vdc
resistive max.
Modbus RTU
128 devices in series max.
(247 devices with repeaters)
Baud rate: 2400, 4800, 9600, or 19200 bps
Fully HART 7 FieldComm compliant. Refer to the
FL500 HART Communication Operating Manual
(PN 10193215).
RX = 50 K CX = 5 nF
FL500 UV/IR Flame Detector