10 User Instructions

To form a reproducible mean value several density measurement are necessary
Remove fat from solvent-resistant sample / /glass sinker /beaker.
Regularly clean sample dishes/glass sinker/beaker, do not touch immersed part
with your hands
Dry sample/glass sinker/pincers after each measurement.
Adjust sample size to sample dish (ideal sample size > 5 g).
Only use distilled water.
When immersing for the first time, lightly shake sample dishes and sinker, in order
to Dissolve air bubbles.
Always ensure that, when re-immersing into the liquid no additional bubbles
adhere; it is better to use pincers to place the sample.
Remove firmly adherent air bubbles with a fine brush or a similar tool.
To avoid adherent air bubbles smoothen samples with rough surface.
Take care that no water drips onto the upper sample dish when weighing with the
help of tweezers.
In order to reduce the surface tension of water and the friction of the liquid on the
wire, add three drops of a common detergent (washing-up liquid) to the
measuring liquid (density modification of dest. water occurring due to the addition
of tensides can be ignored).
Oval samples can be held more easily with pincers when you cut grooves into
them.
The density of porous solids may only be determined approximately. Buoyancy
errors occur when not all the air is eliminated from the pores during immersion in
the measuring fluid.
To avoid great vibrations of the balance, place sample carefully.
Avoid static charge, e. g. dry glass sinker with cotton cloth only.
If the density of your solid only deviates slightly from that of distilled water,
ethanol may be used as measuring fluid. However, check beforehand whether the
sample is solvent-proof. In addition you must observe the applicable safety
regulations when working with ethanol.
Handle glass sinker with care (no warranty claims in case of damage).
To avoid corrosion, don't leave the density set immersed in liquid for a long time.