Section 7 Troubleshooting - Hach SIGMATAX 2 User Manual

Table of Contents

Advertisement

Section 7
Error message
delivery error
filling error
Sensor error
Troubleshooting
As soon as a fault leads to an interruption in operation, SIGMATAX 2 immediately
switches to service mode. Corresponding error messages are shown both on the
SIGMATAX 2 display and on the displays of the connected process photometers.
Service mode means that:
The instrument is no longer in operational mode
The sampling probe and sample storage vessel are empty
The system is depressurized
The sampling probe is closed
The fault indication contact is held in the last valid position
The instrument will not return automatically to operational mode
Possible cause
Sample line clogged
Probe valve clogged
Internal electrical cause
Measures
Select [clean.samp. tube]and carry out a full clean; if
necessary, blow air through the sample line using the
compressor. Inspect the upper level indication sensor.
Check the compressor in the [+FUNCTIONTEST]
menu
Remove probe, extract blockages and inspect the
probe valve and drain valve.
Call customer service.
37