Installation
3.2.2
Connect the fiber optic cable
Figure 13
Fiber optic cable connection
Fiber optic cable for SIGMATAX 2 with PG screw
1
connection
Blind plug for fiber optic cable PG screw connection
2
(PHOSPHAX sigma or TOCTAX)
3.3
Internal tubing
24
1. Carefully screw the level indicator tube (1) and O-ring (2) into the cover (3) from
below.
2. Connect the sampling line (5) to the sampling pipe (4). Then insert the sampling pipe
all the way into the cover through opening 3 or 4.
Note: Fragile glass components
3. Lubricate the rubber seal of the ultrasound transducer (9) with Baysilone paste.
4. Attach the ultrasound transducer (9) to the sample storage vessel using the two
cross-head screws (10). Install the ultrasound transducer in line with the illustration
(cable "upward").
5. Connect the pressure sensor line (7) and insert the sample delivery line (8) 1 cm
(0.39 in) through the central opening in the cover.
Knurled-head screw for opening the front panel
3
Fiber optic cable connection
4