Maintenance
6.1

Maintenance tasks

6.1.1
After 1–4 weeks (as necessary)
Display text
clean. samp. tube
clean. samp. tube
cleaning glass inserted?
sample tube advanced?
opening probe cleaning
emptying probe cleaning
affecting
tube xx sec
please wait xx sec
sample glass inserted?
sample tube retracted
from cleaning to
Test/maint
Operation
opening probe cleaning
emptying probe cleaning
filling probe cleaning
closing probe cleaning
deliv. sample cleaning
emptying sample tube cleaning
[+SERVICEMENU] or operation
34
Potential dangers with contact with chemical/biological substances.
Working with chemical samples, standards and reagents is linked with dangers.
Make yourself familiar with the necessary safety procedures and the correct handling of the
chemicals before the work and read and follow all relevant safety data sheets.
Before handling these substances observe all danger notes and safety information printed on the
containers of the original solutions and in the safety data sheet.
Dispose of all consumed solutions in accordance with the national regulations and laws.
Select the type of protective equipment suitable to the concentration and quantity of the
dangerous material being used.
Clean the sample tube
1. In the [+SERVICEMENU] menu, open the [+INSPECTION] submenu and the
[clean.samp. tube]menu item.
2. The individual steps are specified here and carried out with the support of the
program (see table).
Description/action
Begin the cleaning tasks supported by the program for the sample tube; start
by pressing F4
Confirmation prompt; confirm by pressing F4
Replace the sample storage vessel with the cleaning vessel, which contains
chlorine bleach (sodium hypochlorite), and confirm by pressing F4.
Feed the sample tube through to the base of the cleaning vessel and confirm
by pressing F4.
The cleaning solution is automatically sucked in for 60 seconds. In the
meantime, the sample storage vessel can be cleaned using a paper towel
and, if necessary, a little window cleaning solution or diluted hydrochloric
acid.
Insert the sample storage vessel again and confirm by pressing F4
Retract the sample tube and confirm by pressing F4
At this point, it is possible to choose whether SIGMATAX calls up the
[+SERVICEMENU] menu and remains in SERVICE mode, or whether it
switches to operational mode.
The program runs through an automatic rinsing function to remove the
residual cleaning solution by repeatedly filling and emptying the system.
Status after rinsing is complete
D A N G E R