Access Service Mode - Hach SIGMATAX 2 User Manual

Table of Contents

Advertisement

Operation
5.3.5

Access service mode

32
1. Press one of the function keys, F1 to F4, for 3 seconds.
2. Press F4 twice to scroll to the [+SERVICEMENU] menu.
3. Press F2 to open this menu.
4. When the [+INSPECTION] or [+FUNCTION TEST] menus are selected, the system
automatically switches to service mode.
Service mode means that:
The instrument is no longer in operational mode
The sampling probe and sample storage vessel are empty
The system is depressurized
The sampling probe is closed
The fault indication contact remains in the last valid position
The instrument will not switch back automatically to operational mode