Balance mechanism digital display calibration

The digital display
(6)
ting) to HI (maximum setting). In the unlikely event of this system requiring calibration, proceed as
follows:
Level the platform and apply centre lock (3).
Turn the balance knob
NOTE:
If more than 15 seconds is allowed to elapse between steps, the system will
shut down and revert to its previous settings.
Press and hold the switch
shows CA. Release the switch (9). The display shows HI.
With the balance knob
display will now show LO.
Turn the balance knob
Press and release switch (9).
If the calibration is successful, the display will now show 00.
If unsuccessful, Er will be displayed. Pressing the switch
onds, will allow the system to revert to its previous settings. Calibration may now be carried
out again.
After calibration, rebalance the head (see Balancing the head).