Reference Conditions: Temperature 23°C ± 5K, 45 to 75% RH; battery voltage 6V; conduc-
tor centered in the probe jaws; DC magnetic field; earth's field; no external AC magnetic field;
no external electrical field; sine wave 45 to 65Hz. In Peak, basic accuracy is based on 1 ms
samples and on a signal from 10 to 500Hz.
3.1

Electrical Specifications

CURRENT (TRMS)
Input Range: 1 to 500A
Typical Accuracy (and Phase Shift) with 500A Current Probe MD313:
25A: 5% of Reading (4°)
100A: 2% of Reading (2°)
500A: 2% of Reading (1.5°)
Typical Accuracy (and Phase Shift) with 1000A Current Probe SD652:
50A: ± 0.9% of Reading (1.5°)
200A: ± 0.5% of Reading (0.5°)
1000A: ± 0.5% of Reading (0.5°)
Resolution:
1 to 9.99A: 10mA
10 to 99.9A: 100mA
100 to 500A: 1A
VOLTAGE (TRMS)
Input Range: 0 to 600V
Accuracy:
1 to 99.9V: 0.5% ± 0.6V
100 to 600V: 0.3% ± 2V
Resolution:
0 to 99.9V: 0.1V, 100 to 600V: 1V
TRMS Power Meter Model 3910
CHAPTER 3

SPECIFICATIONS