Calcium Hardness Test Procedure
CA5
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: CA5
3
Press and re-press the MENU button until the display shows the parameter CA5.
FILL METER WITH SAMPLE
4
Rinse the CELL 2 or 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml)
with the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip when the time displays 1. The cursor will
move across the display, informing you that it is about to measure the sample. Record result
displayed (this result is automatically stored in CA5). After testing is completed, rinse cell
immediately.
CA5: Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate, directly
proportional to the Calcium concentration. Values are reported as Calcium Carbonate.
CU6
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: CU6
3
Press and re-press the MENU button until the display shows the parameter CU6.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing
minimizes the potential for cross-contamination from a previous test. Finally, fill cell to
capacity (4ml) with the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip when the time displays 1. The display will
flash "SIt" and begin immediately counting up from 1 to 20 (this extra time allows more thorough
color development). At 20 sec, the cursor will move across the display, informing you that it is
about to measure the sample. Record result displayed (this result is automatically stored in CU6).
After testing is completed, rinse cell immediately.
CU6: Copper reacts with Biquinoline or Bicinchoninic Acid as it is released from the strip to form a purple color,
directly proportional to the copper concentration.
Strip Micro CA, Part No. 486629 from the bottle before
®
Strip Micro CA, Part No. 486629 into the CELL and immediately press READ.
®
Copper Test Procedure
Strip Micro CU, Part No. 486632 from the bottle before
®
Strip Micro CU, Part No. 486632 into the CELL and immediately press READ.
®
7
CA
CU