TR7
TURN METER ON
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: TR7
2
Press and re-press the MENU button until the display shows the parameter TR7.
FILL METER WITH SAMPLE
3
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml)
with the water sample.
ZERO METER
4
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Sample is ready for testing.
DIP STRIP AND PRESS "READ"
5
Dip the eXact
®
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip after "1" on the display disappers. The cursor
will move across the display, while the meter measures the sample (ignore this result). Timer the
reaction in the cell for 100 seconds (timer not included). Press READ again. This starts another
20 SECOND countdown timer. The cursor will move across the display while the meter
measures the sample. Record result displayed (this result is automatically stored in TR7).
USE TABLE
6
Find the "TR7" result in the table below to determine the Phosphate concentration in ppm (parts per
million). (Example: a "TR7" result of 85.3 (use only the 85 for the chart) equals a Phosphate value of
0.39 ppm). Record result. After testing is completed, rinse cell immediately.
Iron results require the table below. Follow eXact
eXact
%T
9
<0.03
90
<0.03
80
0.27
0.30
70
0.60
0.63
60
0.96
1.00
50
1.34
1.40
40
1.80
1.84
30
2.33
2.38
3.02
3.11
20
>4.00
>4.00
10
>4.00
>4.00
0
Phosphate Test Procedure
, Part No. 486814 into the CELL and immediately press READ.
Strip Micro PO
4
Phosphate Table
using eXact
®
Strip Micro PO
®
8
7
6
0.03
0.05
0.33
0.37
0.67
0.70
1.03
1.07
1.43
1.47
1.89
1.95
2.46
2.51
3.20
3.30
>4.00
>4.00
>4.00
>4.00
Micro 7+ Phosphate Test Procedure (above)
®
Strip Micro PO
, Part No. 486814
4
, Part No. 486814 - for 4mL Samples
4
5
4
0.08
0.12
0.39
0.43
0.74
0.78
1.14
1.11
1.53
1.56
2.00
2.05
2.58
2.64
3.40
3.51
>4.00
>4.00
>4.00
>4.00
11
3
2
0.14
0.18
0.21
0.47
0.50
0.54
0.71
0.85
0.89
1.18
1.23
1.27
1.65
1.71
1.62
2.11
2.16
2.22
2.71
2.78
2.86
3.64
3.79
4.00
>4.00
>4.00
>4.00
>4.00
>4.00
>4.00
Rev. 042210 PO4
PO
4
1
0
0.25
0.56
0.92
1.31
1.74
2.27
2.95
>4.00
>4.00
>4.00
Need help?
Do you have a question about the Micro 7+ and is the answer not in the manual?
Questions and answers