Write locking

Write locking prevents any changes to parameters, either via PDM or the handheld
programmer, but still allows access to the device.
2457 (unlock value)
Values
1.
Go to Write Locking:
I
1.
dentification Menu
1. 3. Configuration
2.
Press Right ARROW
3.
Key in 2457 (unlock value) to enable parameter changes.
4.
Press Right ARROW
mode.

Remote operation enable

Remote Operation can be enabled or disabled via the handheld programmer.
Values
To change the setting:
Open Identification Menu, then scroll down to CONFIG.
Press Right ARROW
1. Identification
1.2. Configuration
To enable programming, set REMLOCK to 0. To disable programming, enter 1.
7ML19985JB02
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
Off
any other value
On
1.3.5. Lock
to open Edit mode.
again, to accept the change and return to Navigation
0
Remote operation enabled.
1
Remote operation disabled.
to open the Config Menu, then scroll down to REMLOCK.
1.2.2. Remote Lock
Enables parameter changes.
Disables parameter changes.