Within Program Mode, the handheld programmer has two sub-modes of operation:
1.
Navigation Mode
Press Mode key
Mode: the rightmost digit of the menu number flashes, and the PROGRAM icon
is not visible.
Press Up/Down arrows to change the mode number.
Press Right/Left arrows once to move to the next/previous menu.
Press Right arrow a second time to change the mode from Navigation to Edit
mode.
2.
Edit Mode
In Edit mode, the PROGRAM icon
Press the number keys to enter parameter data.
Press Right arrow to accept the new value and return to Navigation Mode.
Only parameter items can be in Edit mode.
In the case of read-only parameters, no change is permitted.
For Quick Access to parameters via the handheld programmer, press Mode key
Note:
to activate PROGRAM mode, then enter the menu number (see
menu structure
on page 49.

Handheld Programmer: key functions in RUN mode

Key
Updates internal enclosure temperature value and displays it in LCD secondary region.
6
Updates echo confidence value and displays it in LCD secondary region.
8
Updates measurement value and displays it in LCD secondary region.
7ML19985JB02
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
to switch from RUN to PROGRAM and enter Navigation
appears and flashes.
Function
Appendix E: LCD
Page 43