Technical specifications

The following table displays the system specifications for the SimpliAmp
Table 1 SimpliAmp
Thermal Cycler technical specifications:
Feature
[1]
Maximum block ramp rate
Maximum sample ramp rate
Temperature accuracy
Temperature range for
protocol run
Temperature non-uniformity
Dimensions
Weight
PCR volume range
Instrument memory
Display interface
Power
VeriFlex
blocks
Ambient humidity
Ambient operating
temperature
[1]
Reaction volume @ 1 µL
[2]
30 seconds after clock start
[3]
Temperature accuracy ±0.5°C and Temperature non-uniformity < 0.75°C
Table 2 Power and communication ports
Port
6
4°C/sec
[1]
3°C/sec
±0.25°C (35 to 99.9°C)
0°C to 100.0°C
[2]
<0.5°C
8.5 kg (18.3 lb)
USB, on-board
8" color TFT LCD
100-240 V, 50-60 Hz Max: 600 W
15% - 80% Relative Humidity, non-condensing
(acceptable range)
15°C to 30°C (acceptable range)
Description
AC power cable port
10/100 Fast Ethernet port for connecting to a network
USB v2.0 port for connecting to an external network drive, jump drive, or other USB storage device
RS232 serial communication port for connecting the probe for the Temperature Verification Kit
USB v2.0 port for connecting to USB-enabled Wi-Fi Card
Thermal Cycler.
• Height: 21 cm (8.27 in)
• Width: 24.0 cm (9.45 in)
• Depth: 46 cm (18.11 in)
• Supported: 10 - 100 µL
• Settable: 1 - 100 µL
• 3 VeriFlex
zones
• Supported 10°C (5°C zone-to-zone)
[3]
• Settable
20°C (10°C zone-to-zone)