Table of Contents

Advertisement

LED Description

6.7.2 Sync Out

General
The Sync OUT signal is synchronous with the Sync IN or with the frame rate (if the Sync In is
not selected). It can be used to synchronize other events with the camera. It is a TTL signal.
The Sync OUT setup is described in §4.4.7
Characteristics
LED Description
User Manual X8400 sc
LED STATUS
OFF
GREEN
ORANGE
Blinking Green
Blinking Orange
Name
Amplitude (V)
Max frequency (Hz)
Impedance
Minimum Pulse width (ns)
Protection
Connector
LED Status
OFF
Blinking GREEN
GREEN
Description
No signal detected. Check connection
and signal levels
Signal is detected and signal voltage is
correct, but signal is continuous
Signal is detected but signal voltage is
incorrect
Signal is detected.
LED is blinking at signal frequency.
Signal voltage is correct
Signal is detected.
LED is blinking at signal frequency.
Signal voltage is incorrect
Value
TTL signal
0/+5 V
Maximum Frame Rate of the camera for a
given detector configuration
High impedance
300 ns
Voltage peaks (500 V / <1 ns)
Overvoltage (15 V)
Reversed polarity
Coaxial BNC Jack
Description
No signal.
Signal is ready to use.
LED is blinking at signal frequency.
Signal is not usable because it
saturates the camera input.
Signal voltage is 5V
continuously.
38

Advertisement

Table of Contents
loading