8-3.
Example of Sensing data
The below tables provide a few basic examples of the data that is read in Flow Sensor
Registers when the device is subject to given flow and temperature. The values listed in the
tables are given under the hypothesis of perfect device calibration (i.e. no offset, no gain error,
etc).
Table14. Temperature Data registers content vs. Temperature value
Address of Registers
TMP_H
D061h
2Bh
2Eh
26h
19
Value of Registers
TMP_L
D062h
HEX
8Dh
2B8Dh
FFh
2EFFh
BBh
26BBh
D6F-PH MEMS Differential pressure Sensor User's Manual (A288)
Converted
Temperature values
DEC
11149
25.0 degC
12031
48.6 degC
9915
-8.0 degC