Technical data
9.13 FLIR A315f 90°
P/N: 61201-1205
Rev.: 35207
General description
The main purpose of the housing on the FLIR A315f is to increase the environmental specification of
the standard FLIR A315 to IP66 without affecting any of the features available in the camera itself.
The built-in FLIR A315 camera has features and functions that make it the natural choice for anyone
who uses PC software to solve problems and for whom 320 × 240 pixel resolution is sufficient. Among
its main features are GigE Vision and GenICam compliance, which makes it plug-and-play when used
with software packages such as IMAQ Vision and Halcon.
Key features:
Encapsulation to IP66.
Affordable.
GigE compliant.
GenICam compliant.
Trigg/synchronization/GPIO.
16-bit 320 × 240 pixel images at 60 Hz, signal, temperature linear, and radiometric.
Compliant with any software that supports GenICam, including National Instruments IMAQ Vision
and Stemmers Common Vision Blox.
Typical applications:
Imaging and optical data
IR resolution
Thermal sensitivity/NETD
Field of view (FOV)
Minimum focus distance
Focal length
Spatial resolution (IFOV)
Lens identification
F-number
Image frequency
Focus
Detector data
Detector type
Spectral range
Detector pitch
Detector time constant
Measurement
Object temperature range
Accuracy
Measurement analysis
Atmospheric transmission correction
Optics transmission correction
Emissivity correction
Reflected apparent temperature correction
#T559794; r. AJ/35709/35709; en-US
320 × 240 pixels
< 0.05°C @ +30°C (+86°F) / 50 mK
90° × 73°
20 mm (0.79 in.)
4 mm (0.157 in.)
6.3 mrad
Automatic
1.3
60 Hz
Automatic or manual (built in motor)
Focal plane array (FPA), uncooled
microbolometer
7.5–13 µm
25 µm
Typical 12 ms
–20 to +120°C (–4 to +248°F)
0 to +350°C (+32 to +662°F)
±4°C (±7.2°F) or ±4% of reading
Automatic, based on inputs for distance, atmos-
pheric temperature and relative humidity
Automatic, based on signals from internal sensors
Variable from 0.01 to 1.0
Automatic, based on input of reflected
temperature
47