Philips ST80i Instructions For Use Manual page 183

Stress test system
Table of Contents

Advertisement

Table A-1
Troubleshooting ST80i issues
Problem
Data-Entry Issues
Cannot manually enter
patient data
Wireless Patient Module Issues
Packet errors/ signal
drop
Wireless Patient Module Issues
No waveform signal on
the screen
Cause
It is possible that the Export
check box has been inadvertently
selected on the
Miscellaneous
in the Configuration settings.
Signal on the screen shows a red
dashed line.
(continued)
With the patient (or simulator)
connected, no waveforms are
displayed; only the red dashed
line and lead-fail indicator appear.
NOTE If neither the signal nor the
red dashed line is present, click the
"X" in the upper right corner of the
screen, and then click OK in the
"Abort Test" dialog box.
Solution
For the affected profile(s), do the following:
HL7
Click
1
Settings
tab
Miscellaneous
Ensure the
2
Export: HL7
Click
3
Save Profile
Configure the system to automatically hop
the RF channel, or perform an RF Scan to
locate the source of the interference and to
change to the optimal channel for the
wireless patient module. For details about the
wireless settings, see the ST80i Wireless
Patient Module Instructions for Use.
Check the ST80i
the wireless patient module to ensure that the
two units are set to the same channel.
For details about setting the receiving
channel settings, see the ST80i Installation
and Configuration Guide.
For details about the wireless settings, see the
ST80i Wireless Patient Module Instructions
for Use for instructions.
ST80i Stress Test System Instructions for Use
Troubleshooting ST80i Issues
, then click the
tab.
check box is clear.
and
.
Exit
settings, as well as
System
A-7