The Patient Session
Global Interpretive Settings
In the Settings section of the ST80i application, you can configure the Global Interpretive
Settings for DXL Algorithm and CALg templates, include the following:
DXL Algorithm
Adult Bradycardia
Borderline Statement Sensitivity
Pacer Detection
Report Confirmation Label
CALg Templates
Arrhythmia Analysis
ST Segment Analysis
J-ST Settings
Global Interpretive Settings also include:
Arrhythmia Event Notification
ST Amplitude Calculation
To configure the Global Interpretive Settings, see the ST80i Installation and Configuration
Guide.
Pace-Pulse Detection
The Pace-Pulse detection algorithm detects both atrial and ventricular pacing. Pacemaker
detection is preset in the user profile. The options include:
Paced
Paced (Magnet)
Non-Paced
Not Known if Paced
"Not known if paced" is the default mode, in which the detector uses its default sensitivity
setting to determine if pacemaker pulses are present.
The pace-pulse detection setting in effect at the time a snapshot is taken is used for the analysis
of that snapshot. If the pace-pulse program is enabled and it detects the presence of pace-
pulses, it will provide output in textual and graphical formats:
The interpretive report will contain textual statements about the findings for the pace-
pulse program.
3-40
ST80i Stress Test System Instructions for Use