AEMC 6255 User Manual page 32

Micro-ohmmeter
Hide thumbs Also See for 6255:
Table of Contents

Advertisement

1.800.561.8187
4.5.4. Low Inductive Resistance Automatic Triggering Measurement
Mode
This mode is intended only for measuring resistance on material without a time
constant (non-inductive). The use of the spring loaded Kelvin probes (Cat.
#2118.77 or Cat. #2118.78) listed in the Accessories section is recommended for
this mode.
From the Stand-by state, start a test by performing the following:
1.
Press the
2.
Press the START/STOP button.
3.
Connect the probes to the specimen. The unit waits until it detects that
the current and voltage leads are connected correctly.
4.
Residual voltage (V
0
measured.
5.
The measurement current (I) is established, the voltage across the
resistor terminals (V
result R = (V
- V
) / I is displayed.
1
0
6.
To start a new measurement cycle, Release at least one probe, and
then reconnect it to the next point or specimen.
If the range is exceeded, the instrument displays "Err 07". Changing
the range switch position will stop the test cycle and return to the
Stand-by state. Each measurement taken during the test cycle can
be temperature compensated and stored while the test cycle is
running.
7.
Store the measurement of each specimen or point by first pressing the
MEM button, then select the object and test location to store the
measurement using the arrow buttons. When the desired location has
been selected, press the MEM button a second time to complete the
data storage process.
8.
The test is stopped by pressing the START/STOP button.
32
www.
button until the AUTO icon shows on the display.
) measurement across the resistor terminals is
) is measured and the measurement resistance
1
Micro-Ohmmeter Model 6255
.com
information@itm.com

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents