Keypad In Run Mode; Keypad In The Setup / Commissioning / Calibration Menu - Novatech 1732 Technical Manual

Oxygen transmitter
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

Button Text
1
S
/ R
ETUP
UN
2
D
/ F
ISPLAY
UNCTION
3
D
/ F
ISPLAY
UNCTION
4
A
/ O
LARM
PTION
5
A
/ O
LARM
PTION
6
A
A
/ E
LARM
CCEPT
NTER
7
G
1 P
1 / S
AS
URGE
ENS
8
G
2 P
2 / A
AS
URGE
UTO

5.3.1 Keypad in Run Mode

After the transmitter is powered on and has completed its initial startup procedure, it will enter the Run Mode.
The button functions in this mode are as follows;
S
/ R
ETUP
UN
By pressing this button once from Run Mode the transmitter will enter the Setup Menu system. In the Setup
Menu the function of each of the buttons is reassigned. Pressing the S
while in the Setup Menu will return the transmitter to the Run Mode. If the transmitter is left idle in the Setup
Menu for more than 60 seconds, it will automatically return to Run Mode.
 / D
D
ISPLAY
ISPLAY
The display buttons are used to scroll the lower line up and down through the variety of measurements that
are available on the lower line. For a complete list of options see chapter 6.4.2 Lower Line Items.
A
LARM
If there is either a new alarm or an active alarm the A
status. The Alarm Light will be flashing if there is a new alarm or steady if there is an existing alarm. For
more details on the A
LARM
show that the transmitter is now in the alarm display mode.
A
LARM
When the transmitter is displaying active alarms (the A
 button allow the operator to examine the date / time of the alarm and the date / time that the
and A
LARM
alarm was acknowledged.
If the transmitter was in Run Mode when this button is pressed it will go into the alarm log display mode. See
Chapter 9. Alarms for more details.
A
A
LARM
CCEPT
Press this button to acknowledge the currently displayed alarm (See Chapter 9. Alarms).
G
1 / P
1 G
AS
URGE
AS
These two buttons are used to turn on the gas / purge solenoids. When the transmitter is in the manual cal /
purge mode (Commissioning Menu functions #23 and 32) the solenoid will be activated for as long as the
button is pressed. When the transmitter is in the auto cal / purge mode the automatic cal / purge cycle is
started. The cycle can be stopped by pressing the same button again.

5.3.2 Keypad in the Setup / Commissioning / Calibration Menu

NOTE: Access to the Commissioning Menu and Calibration Menu can be disabled by switching OFF the
corresponding slider on DIP switch SW1 on the 1730 Main PCB. To enter either Commissioning Menu or
Calibration Menu, the corresponding DIP switch SW1 on the 1730-1 PCB must be turned on.
August 2018
1732 Oxygen Transmitter
Run Mode (white text)
Enter Setup Menu
Display next Lower Line item
Display last Lower Line Item
Next alarm
Previous alarm
Acknowledge displayed alarm
Gas 1 / Purge 1 manual
I
MP
activate
Gas 2 / Purge 2 manual
C
AL
activate
M
and buttons see Chapter 9. Alarms. The Setup Light will flash slowly to
ODE
2 / P
2
URGE
Setup Menu (black/blue text)
Return to Run Mode
Next function
Previous function
Next / increment option
Previous / decrement option
Save current option
Check Probe impedance
Manually perform device
calibration
/ R
ETUP
 button can be pressed to examine the alarm
LARM
 button has been pressed), the A
LARM
button a second time
UN
 button
LARM
Technical Manual