4.3 Installing a 1234 Oxygen Sensor

Mounting - Screw the 1234 Oxygen Sensor to a wall or similar surface with the piping connections at the
top.
Sample Piping - Connect the gas sample piping to the "sample in" port. If the process, boiler, kiln or furnace
has a positive pressure, no suction will be required. If the sample is under a negative pressure, connect a
pump to the "inlet" port. The flow rate should be within the range of 1 to 5 litres/minute (2 to 10 scfh).
Upward sloping
sample line 3/8"
Flue
stainless steel
Gas
Sampling
Probe
Flowmeter

4.4 Installing the Auxiliary Thermocouple

Weld a 1/2-inch BSP mounting socket to the flue within about 300 mm (12"), and upstream of the oxygen
probe. The thermocouple should be of similar length to the oxygen probe to prevent flue temperature
distribution errors.
The thermocouple should be connected to terminals 7 & 8. These terminals will not be available for an
auxiliary thermocouple if a second probe has been installed.