Display And Keypad; Run Mode Display - Novatech 1732 Technical Manual

Oxygen transmitter
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

The 1732 Oxygen Transmitter has a graphic display, 8 buttons and 5 LED indicators to show the status of
the transmitter.
All of the buttons have multiple functions, depending on what is currently on the display. As a general
starting point, the larger white text on the button is the function while the transmitter is in the Run Menu and
the smaller black text on the button is the function in the Setup / Commissioning / Calibration Menus.
When the transmitter is sitting idle it will revert to Run Mode, in which standard information relating to the
process appears on the screen. To configure the transmitter, the operator must access Setup Menu. This is
achieved by pressing, and in some cases holding down the S
transmitter will return to the Run Mode when the S
seconds of inactivity.

5.1 Run Mode Display

In Run Mode the 1732 shows the first oxygen probe measurement in large characters at the top of the
display and a user selectable lower line in smaller characters below. Other items on the display include the
activity heartbeat indicator in the top right corner, a row of single letter action indicators in the bottom left
corner and the current time in the bottom right corner.
Power Indicator
Probe 1 Oxygen
Lower line secondary
process variable
August 2018
1732 Oxygen Transmitter
ETUP
The front panel of the 1732 Oxygen Transmitter
12.25%
Probe 1 Temp 720.3 C
B
5. D
ISPLAY AND
button (see Chapters 6, 7 8). The
ETUP
button is pressed again or after a period of 60
14:20:36
Activity Information Indicators:
B Burner input enabled (terminals 11&12)
A The transmitter is performing an automatic
ADC calibration.
T (flashing) One or more oxygen probes is
below operating temperature (650°C / 1200°F)
Z The transmitter is performing a probe
impedance check.
K
EYPAD
Heartbeat Indicator
Current time
Technical Manual
29