Setup Menu; Function Summary Table; Setup Menu Display; Changing Menu Options - Novatech 1732 Technical Manual

Oxygen transmitter
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

This chapter describes the functions available when the Setup Menu is entered on the transmitter.
The Setup Menu is mode is accessed from Run Mode by pressing the S
words 'Setup Menu' appear at the bottom of the display. The transmitter will return to the Run Mode when
the S
button is pressed again or after 60 seconds of keypad inactivity.
ETUP

6.1 Function Summary Table

The following table shows a summary of Setup Menu functions:
Menu #
Function name
01
Probe 1 Offset
02
Probe 2 Offset
03
Lower Line Items
04
Oxygen Display Units
05
Damping Factor
06
Process Alarms

6.2 Setup Menu Display

Power indicator
Function Number
Selected Option
Activity Indicator:
A Autocal active
Z Probe impedance
check active

6.3 Changing Menu Options

The purpose of having an interactive Setup Menu is to allow for configuration of the transmitter using the
graphical display and keypad.
Once an option is changed and entered using the E
device configuration and calibration is stored into the non-volatile memory and will be retained permanently
even if the device does not have power.
To reset the transmitter configuration to factory default settings, see chapter 10.2 Cold Start.
To change an option in the Setup Menu system:
1. Enter the Setup Menu by pressing the S
the display will have the format shown below. The operations of the buttons are now the operations
written in WHITE on the keypad. The menu name is written at the bottom of the display.
2. While in the Setup Menu the required function can be selected by using the F
 buttons. The options available for that function can be seen by using the O
F
UNCTION
 buttons.
O
PTION
3. When the required option is on the display press the E
When finished, press the S
August 2018
1732 Oxygen Transmitter
Range
-6.0 to +6.0mV
-6.0 to +6.0mV
see chapter 6.4.2 Lower Line Items
Oxygen % / Oxygen ppm
No Damping to 10 Samples
Enabled / Disabled
01
Probe 1 Offset
+0.0 mV
A

Setup Menu

ETUP
/ R
button to return to the Run Mode.
ETUP
UN
ETUP
Saved
button that value immediately becomes active. The
NTER
/ R
button once. The Setup Light will come on and
UN
button to save that value.
NTER
6. S
M
ETUP
button momentarily so the
Default value
0.0mV
0.0mV
Oxygen %
5 Samples
Disabled
Function Name
Verification that the
selected option has
been saved
Menu name
 and
UNCTION
PTION
Technical Manual
ENU
 and
33