Device Specifications; Hardware Specifications; Transmitter Specifications - Novatech 1732 Technical Manual

Oxygen transmitter
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

3.1 Hardware Specifications

3.1.1 Transmitter Specifications

Number of Oxygen Probes:
Oxygen Range:
Oxygen Accuracy:
Thermocouple Types:
Temperature Accuracy:
Analog Outputs:
(WARNING: DO NOT LOOP POWER OUTPUTS. Use only passive receivers for commissioning and
testing. The use of loop powered receivers will damage the output)
Output Load:
Alarm Relays:
Alarm Relay Contacts:
(WARNING: Do not use both mains voltage and low voltage connections to adjacent alarm contacts)
Mains Voltage Supply:
Overvoltage:
Power:
Environmental Rating:
Altitude
Degree of Protection:
Case Size:
Case Weight:
WARNING: All signal level connections onto the transmitter must be treated as safety extra-low
voltage (SELV) as defined in the standard IEC61140. Double insulation must be used when
connecting these terminals to systems that might carry high voltage.
August 2018
1732 Oxygen Transmitter
1 or 2
-30
1 x 10
to 100%
±1% of actual measured oxygen value with a repeatability of ±0.5% of the
measured value
Type K and J
±2C
0-20mA or 4-20mA field selectable
Active Outputs
1000 ohm max
4
2A 240VAC, 2A 30VDC
Reinforced insulation when used with mains voltage
100 to 240VAC 50/60 Hz
Reinforced insulation
Category II (IEC60364-4-443)
5 Watts for controller plus probe power
530W max., 25% duty cycle each probe on 240VAC
110W max., 100% duty cycle each probe on 110VAC
576W (2.4A) max
Operating Temperature -25°C to 55°C
Relative Humidity 5% to 95% (non-condensing)
2000m Maximum
IP65
IP54 with internal reference air pump
315mm (12.4") wide, 190mm (7.5") high, 110mm (4.3") deep
3.3 Kg (7.3 lbs.)
3. D
S
EVICE
PECIFICATIONS
Technical Manual
11