Calibration Menu; Function Summary Table - Novatech 1732 Technical Manual

Oxygen transmitter
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

This chapter describes the functions available in the Calibration Menu on the transmitter. For specific
information about calibrating the transmitter see Chapter 10. Instrument Calibration.
To Access the Calibration Menu;
Ensure the DIP switch SW1-3 on the main PCB is in the ON position
From Run Mode, press and hold the Setup button for approximately 4 seconds until the words
"Calibration Menu" appears at the bottom of the display.
When the transmitter is in the Calibration Menu the Setup Light will be on and the words "Commissioning
Menu" will be shown at the bottom of the display. The transmitter will return to the Run Mode when the
S
button is pressed again or after 60 seconds of keypad inactivity.
ETUP
Changing options in the Calibration Menu is the same as the Setup Menu. see chapter 6.3 Changing Menu
Options.

8.1 Function Summary Table

The following table shows a summary of Calibration Menu functions:
Menu #
Function name
01
Reference Voltage 1, 50mV
02
Reference Voltage 2, 200mV
03
Reference Voltage 3, 1200mV
04
Reference Voltage 4, 2500mV
05
Output Channel 1, Calibration
06
Output Channel 1, 4mA Trim
07
Output Channel 1, 20mA Trim
08
Output Channel 2, Calibration
09
Output Channel 2, 4mA Trim
10
Output Channel 2, 20mA Trim
Ambient Temperature Sensor
11
Offset
12
Low Oxygen Cal 1
13
Low Oxygen Cal 2
14
Ref Gas Probe 1
15
Ref Gas Probe 2
16
Transmitter Output Select
17
Transmitter Output Limiting
Mains Voltage Detection
18
Override
Mains Frequency Detection
19
Override
20
Heater SSR Selection
21
SSR Fail Protection
22
Burner Temp O/Ride
23
Ref Pump Cycling
24
Dry Reference Air
August 2018
1732 Oxygen Transmitter
8. C
Range
40.00 to 60.00mV
180.00 to 210.00mV
1150.0 to 1250.0mV
2400.0 to 2550.0mV
Auto Calibrated / Manual Calibrated /
Set 4mA / Set 20mA
3.00 to 5.00mA
19.00 to 21.00mA
Auto Calibrated / Manual Calibrated /
Set 4mA / Set 20mA
3.00 to 5.00mA
19.00 to 21.00mA
-10.0°C to 10.0°C
80.0% to 120.0%
Ambient Air, 0.1% to 25%
4-20mA / 0-20mA / 4-20mA over-
scale limit / 0-20mA over-scale limit
Disabled / Hold 0mA / Hold 4mA /
Hold 20mA
Automatic /
220/240 / 110/120
Automatic /
50Hz / 60Hz
Normal
Heater1 <-> Heater2
Heaters <-> CalPurge
Enabled / Disabled
Enabled / Disabled
Always On / Cycle Above xx°C
(where xx is ranged 30-55°C)
Wet Air / Dry Ambient / Dry Air
M
ALIBRATION
Default value
47.14mV
182.24mV
1221.8mV
2489.2mV
Auto Calibrated
4.00mA
20.00mA
Auto Calibrated
4.00mA
20.00mA
0.0°C
100.0%
Ambient Air
4-20mA
Hold 0mA
Automatic
Automatic
Normal
Enabled
Disabled
Cycle Above 35°C
Wet Air
Technical Manual
ENU