4.3 Keypad

There are 8 buttons built into the label on front panel of the 1732 Oxygen Transmitter. The button functions
are written in BLACK and WHITE to identify the function of the button in either Run Mode or the Setup Menu
system.
Setup Light
1
Button Text
1
S
/ R
ETUP
UN
2
D
/ F
ISPLAY
UNCTION
3
D
/ F
ISPLAY
UNCTION
4
A
/ O
LARM
PTION
5
A
/ O
LARM
PTION
6
A
A
/ E
LARM
CCEPT
NTER
7
G
1 P
1 / S
AS
URGE
ENS
8
G
2 P
2 / A
AS
URGE
UTO

4.3.1 Keypad in Run Mode

After the transmitter is powered on and has completed the initial startup procedure it will enter the Run Mode.
In this mode the top line of the display will show the oxygen measurement from probe 1. The other button
functions are –
S
/ R
ETUP
UN
By pressing this button once from Run Mode the transmitter will enter the Setup Menu system. In the Setup
Menu the function of each of the buttons is reassigned. Pressing the S
while in the Setup Menu will return the transmitter to the Run Mode. If the transmitter is left idle in the Setup
Menu for more than 60 seconds it will automatically return to Run Mode.
∆ / D
D
ISPLAY
ISPLAY
The display buttons are used to scroll the lower line up and down through the variety of measurements that
are available on the lower line. For a complete list of options see chapter 5.3.2 Lower Line Items.
A
LARM
If there is either a new alarm or an active alarm the A
status. The Alarm Light will be flashing if there is a new alarm or steady if there is an existing alarm. (see
Chapter 6. Alarms). The Setup Light will flash slowly to show that the transmitter is now in the alarm display
mode.
February 2015
1732 Oxygen Transmitter
2
3
Run Mode (white text)
Enter Setup Menu
D
ISPLAY NEXT
Display last Lower line Item
Next alarm
Previous alarm
Acknowledge displayed alarm
Gas 1 / Purge 1 manual
I
MP
activate
Gas 2 / Purge 2 manual
C
AL
activate
Alarm Light
4
6
5
Setup Menu (black/blue text)
Return to Run Mode
L
N
OWER LINE ITEM
EXT FUNCTION
Previous function
Next / increment option
Previous / decrement option
Save current option
Check Probe impedance
Manually perform device
calibration
∆ button can be pressed to examine the alarm
LARM
Cal / Purge Lights
7
/ R
button a second time
ETUP
UN
Operators Manual