Commissioning Menu; Function Summary Table - Novatech 1732 Technical Manual

Oxygen transmitter
Hide thumbs Also See for 1732:
Table of Contents

Advertisement

This chapter describes the functions available in the Commissioning Menu on the 1732 Oxygen Transmitter.
To Access the Commissioning Menu;
Ensure the DIP switch SW1-2 on the main PCB is in the ON position
From Run Mode, press and hold the Setup button for approximately 2 seconds until the words
"Commissioning Menu" appears at the bottom of the display.
When the transmitter is in the Commissioning Menu the Setup Light will be on and the words
"Commissioning Menu" will be shown at the bottom of the display. The transmitter will return to the Run
Mode when the S
button is pressed again or after 60 seconds of keypad inactivity.
ETUP
Changing options in the Commissioning Menu is the same as the Setup Menu. See chapter 6.3 Changing
Menu Options.

7.1 Function Summary Table

The following table shows a summary of Commissioning Menu functions:
Menu #
Function name
01
Internal Clock Date
02
Internal Clock Time
03
Service Date
04
Number of Probes
05
Probe 1 Type
06
Probe 2 Type
07
Probe 1 TC Type
08
Probe 2 TC Type
09
Auxiliary TC Type
10
Transmitter Output Channel 1
11
Transmitter Zero Channel 1
12
Transmitter Span Channel 1
13
Transmitter Output Channel 1
14
Transmitter Zero Channel 2
15
Transmitter Span Channel 2
16
Flue Pressure Units
17
Flow Pressure Input
18
Flue Pressure Value (Fixed)
19
Flue Input Zero
20
Flue Input Span
21
Temperature Units
22
Cal Freezes Outputs
23
Solenoid 1 Operation
24
Solenoid 1 Auto / Manual
25
Solenoid 1 Start Time
26
Solenoid 1 Period
27
Solenoid 1 Duration
28
Solenoid 1 Post Freeze
29
Oxygen Content Cal Gas 1
30
Max Cal Gas 1 Positive Error
August 2018
1732 Oxygen Transmitter
7. C
OMMISSIONING
Range
Single / Dual Probe
1231 / 1234 Heated
or 1232 Unheated
K, J, R or S Types
K, J, R or S Types
K, J, R or S Types, Disabled
See chapter 7.2.7 Transmitter
Output Channel
Inches WG, mm WG, kPa, PSI
Fixed / Variable
-1 to 3 Atm
Celsius / Fahrenheit
Enabled / Disabled
Calibration Gas / Purge
Automatic / Manual
00:00 to 23:45
1 minute – 7 days
1 – 90 seconds
5 – 300 seconds
0.1 – 20.9%
0.1% to 3.0%
M
ENU
Default value
Single Probe
1231 / 1234 Heated
1231 / 1234 Heated
K-Type
K-Type
Disabled
Linear oxygen
Linear oxygen
Inches WG
Fixed
0 Inches WG
0 Inches WG
100 Inches WG
Celsius
Enabled
Calibration Gas
Manual
00:00 (midnight)
1 hour
30 seconds
60 seconds
8.0%
0.5%
Technical Manual
37