4.5a Shield Connections

All external wiring to the 1732 transmitter should be shielded. Do not connect shields at the field end.
Simply clip off and insulate. There are two M4 earth screw terminals available in the 1732 transmitter. An
extra terminal strip may be required to connect all shields together. This should be supplied by the installer.

4.5b Earth connection (PE)

The PE (protective earth) input connection must be made to the earth stud on the right hand side of cabinet.
The PE input connection must be the first connection onto the earth stud and it must be secured by a
separate nut and spring washer. All other earth connections (bonding connections) can be made on either of
the two earth studs in the base of the cabinet.
The transmitter MUST be securely earthed.
Incoming
PE wire
Assembly of the PE and bonding connections on the earth stud