Memory Devices - Tektronix MDO4104C Declassification And Security Instructions

Mdo4000c series
Table of Contents

Advertisement

Clear and Sanitize Procedures

Memory Devices

Table 1: Volatile memory devices
Type and minimum
size
Function
CMOS, SDRAM,
Contains
128 Mbit x 16, 8 devices,
waveform
2 Gb ea, 16 Gb total,
main
DDR2, 1.8 V, 400 MHz
data
(Effective 800 MHz);
during
MT47H128M16RT-25E,
operation
BGA
1
CMOS, SDRAM,
Contains
64 Mbit x 16, 8 devices
waveform
1 Gb ea, 8 Gb total,
main
DDR2, 1.8 V, 400 MHz
data
(Effective 800 MHz);
during
MT47H64M16HR-25E
operation
BGA
2
CMOS, SDRAM,
Contains
64M x 16, 1024 Mb
instrument
DDR2, 1.8 V, 400 MHz
code
(Effective 800 MHz);
and
MT47H64M16HR-25E
data
during
3
operation
CMOS, SDRAM,
Contains
64M x 16, 1024 Mb
display
DDR2, 1.8 V, 400 MHz
images
(Effective 800 MHz);
during
MT47H64M16HR-25E
operation
4
CMOS, DRAM;
Contains
4M X 32,128 Mb,
code for
SDRAM, 3.3 V;
digital
MT48LC4M32B2B\5-7,
waveform
90-BALL FBGA
processor
(11 MM X 13 MM),
DS4
2
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options. Detailed procedures to clear or sanitize
these devices, if any, are shown following each table.
Type
Method
of user
Backed
of
info
up by
modifi-
stored
battery
cation
Both
No
Indirect
Both
No
Indirect
Both
No
Indirect
Both
No
Indirect
None
No
Indirect
Data input
method
Location
Front panel control
Main
and oscilloscope
Acquisition
input channels
board
Front panel control
Main
and oscilloscope
Acquisition
input channels
board
Front panel control
Main
and oscilloscope
acquisition
input channels.
board
Data inputs through
USB port and
Ethernet port.
Front panel control
Main
and oscilloscope
acquisition
input channels.
board
System
Main
acquisition
board
Declassification and Security Instructions
User
To clear or
accessible
sanitize
No
Remove
power
from the
instrument
for at least
20 seconds
No
Remove
power
from the
instrument
for at least
20 seconds
No
Remove
power
from the
instrument
for at least
20 seconds
No
Remove
power
from the
instrument
for at least
20 seconds
No
Remove
power
from the
instrument
for at least
20 seconds