  • Page 1 MDO3000 Series Mixed Domain Oscilloscopes Service Manual *P077098100* 077-0981-00...
  • Page 3 MDO3000 Series Mixed Domain Oscilloscopes Service Manual Revision A Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
  • Page 4 Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
  • Page 5: Table Of Contents

    Front-Panel Board ..................... Adjustments ......................Maintenance......................Preventing ESD ....................Inspection and Cleaning..................Module Removal....................Troubleshooting....................Troubleshooting Procedure .................. Unpacking and Repacking Instructions ..............Replaceable Parts List..................... Parts Ordering Information .................. Using the Replaceable Parts List................MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 6 Table of Contents List of Figures Figure 1: MDO3000 Series block diagram ..............Figure 2: Removing tabs from the front protective cover............. Figure 3: Module locator ..................Figure 4: Primary troubleshooting procedure ..............Figure 5: AC power supply troubleshooting procedure ............
  • Page 7 Table 2: Internal inspection checklist ................Table 3: Parts list column descriptions................Table 4: Front panel assembly ................... Table 5: Display, analog board, and main board ............... Table 6: Rear chassis and connecting modules ..............MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 8: Important Safety Information

    To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, make sure that the product is properly grounded. Do not disable the power cord grounding connection. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 9 Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product. Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings and markings on the product.
  • Page 10 Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries. Use only the Tektronix rackmount hardware specified for this product. Probes and test leads Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
  • Page 11: Service Safety Summary

    Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components. Verify safety after repair. Always recheck ground continuity and mains dielectric strength after performing a repair. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 12: Terms In This Manual

    find out the nature of the potential hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.) The following symbol(s) may appear on the product: viii MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 13: Preface

    Preface This service manual provides information that you need to troubleshoot, disassemble, and replace parts on the following Tektronix oscilloscopes: Model Bandwidth Analog channels RF frequency 100 MHz MDO3012 Up to 3 GHz MDO3014 100 MHz Up to 3 GHz...
  • Page 14 Preface MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 15: Where To Find Operating Information

    Where to Find Operating Information For information on installing and operating your oscilloscope, refer to the Tektronix MDO3000 Series Mixed Domain Oscilloscopes User Manual, which was included on the documentation CD provided with your oscilloscope. This manual is available in 11 languages.
  • Page 16 Where to Find Operating Information MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 17: Theory Of Operation

    The Display system combines live waveform data from acquisition memory with menus and text, and stores this information in display memory. It then uses this data to refresh the WVGA display module (LCD). MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 18: Front-Panel Board

    Main board, and controls power to the system. The Front Panel board also generates the probe compensation output signal, provides an interface to the application modules, and controls power to the system. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 19: Figure 1: Mdo3000 Series Block Diagram

    Theory of Operation Figure 1: MDO3000 Series block diagram MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 20 Theory of Operation MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 21: Adjustments

    Adjustments Please return the instrument to Tektronix if adjustment is required. There are no adjustment procedures (calibrations) that can be performed by the customer. Calibrations can only be performed at an authorized Tektronix Service Center and require the use of traceable test equipment (signal sources and measuring equipment) to verify the performance of the instrument.
  • Page 22 Adjustments MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 23: Maintenance

    Inspection and cleaning are done as preventive maintenance. Preventive maintenance, when done regularly, may prevent oscilloscope malfunction and enhance its reliability. Preventive maintenance consists of visually inspecting and cleaning the oscilloscope and using general care when operating it. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 24 If any dirt remains, use a cloth or swab dipped in a 75% isopropyl alcohol solution. Use a swab to clean narrow spaces around controls and connectors. Do not use abrasive compounds on any part of the chassis that might damage the chassis. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 25: Table 1: External Inspection Checklist

    If any circuit board is repaired or replaced, adjustment is required. (See page 7, Adjustments.) CAUTION. To prevent damage from electrical arcing, make sure that circuit boards and components are dry before applying power to the oscilloscope. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 26: Table 2: Internal Inspection Checklist

    Dry all parts with low-pressure, deionized air. d. Dry all components and assemblies in an oven or drying compartment using low-temperature (125 °F to 150 °F) circulating air. Lubrication There is no periodic lubrication required for this oscilloscope. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 27: Module Removal

    Failure to do so could cause serious injury or death. Required Equipment You will need a Torx driver with a T-15 Torx tip to remove instrument screws. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 28: Figure 3: Module Locator

    fingers. The oils on your fingers will degrade or damage the switch contacts. To help prevent damage to the keypad, use cotton gloves when removing or installing the keyboard pad. Figure 3: Module locator MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 29: Troubleshooting

    When you operate the instrument with the cabinet removed, the system fan will not operate. You must provide an external fan to cool the interior of the instrument. Failure to do so could cause instrument damage. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 30: Figure 4: Primary Troubleshooting Procedure

    Maintenance Figure 4: Primary troubleshooting procedure MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 31: Figure 5: Ac Power Supply Troubleshooting Procedure

    Maintenance Figure 5: AC power supply troubleshooting procedure NOTE. The test point voltages are printed on the Main board. The +12 V test point is on the upper right-hand corner of the board. MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 32: Figure 6: Module Isolation Troubleshooting Procedure

    Maintenance Figure 6: Module isolation troubleshooting procedure MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 33: Unpacking And Repacking Instructions

    A complete description of the service required Seal the shipping carton with an industrial stapler or strapping tape. Mark the address of the Tektronix Service Center and also your own return address on the shipping carton in two prominent locations. See www.tektronix.com/service to find a service center near you.
  • Page 34 Maintenance MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 35: Replaceable Parts List

    Level revision level number. When you order parts, Tektronix will provide you with the most current part for your product type, serial number, and modification (if applicable). At the time of your order, Tektronix will determine the part number revision level needed for your product, based on the information you provide.
  • Page 36 Module Servicing Modules can be serviced by selecting one of the next three options. Contact your local Tektronix service center or representative for repair assistance. Module exchange. In some cases you may exchange your module for a remanufactured module. These modules cost significantly less than new modules and meet the same factory specifications.
  • Page 37: Using The Replaceable Parts List

    Items in this section are referenced by figure and index numbers to the exploded view Number illustrations that follow the list. Tektronix Part Use this part number when ordering replacement parts from Tektronix. Number 3 and 4 Serial Number Column three indicates the serial number at which the part was first effective. Column four indicates the serial number at which the part was discontinued.
  • Page 38: Table 4: Front Panel Assembly

    5 screws across the bottom of the Front panel assembly attach it to the front case assembly – without contact springs. 260-2889-00 SWITCH, KEYPAD, ELASTOMERIC, UTILITY 407-2567-00 BRACKET, SUPPORT, UTILITY SWITCH 260-2971-00 SWITCH, KEYPAD, ELASTOMERIC, FRONT PANEL MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 39 GUIDE, KEY; POLY, BAYBLEND FR-110 213-1149-00 SCREW, TPG, TF; 2-28 X.5, PLASTITE, FLAT HEAD, PHILLIPS, STL, ZNPL 259-0212-00 CIRCUIT, FLEX; 10 KEY 200-5301-00 COVER; OPTION KEY DOOR, PC/ABS, FR110, SILVER GRAY 200-5052-00 COVER, FRONT PROTECTIVE MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 40: Figure 7: Exploded View, Front Panel Assembly

    Replaceable Parts List Figure 7: Exploded view, front panel assembly MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 41: Table 5: Display, Analog Board, And Main Board

    The Main board and the Attenuator board must be calibrated together if either is found defective. 878-0803-xx ATTENUATOR BD, 1 GHz, 2 CH. Return the instrument to Tektronix for service. The Main board and the Attenuator board must be calibrated together if either is found defective. 878-0802-xx ATTENUATOR BD, 1 GHz, 4 CH.
  • Page 42: Figure 8: Display, Analog Board, And Main Board Assembly

    Replaceable Parts List Figure 8: Display, analog board, and main board assembly MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 43: Table 6: Rear Chassis And Connecting Modules

    CABLE CLAMP 131-6643-00 CONTACT, ELEC: GROUNDING, ELECTROLESS NICKEL PLATE 878-0761-xx I/O BOARD ASSEMBLY 119-8180-00 SYSTEM FAN ASSEMBLY, 120 MM, TUBEAXIAL, 12 VDC; SAFETY CONTROLLED, WITH BRACKET 119-8042-00 POWER SUPPLY; ROHS COMPLIANT; SAFETY CONTROLLED MDO3000 Series Mixed Domain Oscilloscopes Service Manual...
  • Page 44: Figure 9: Rear Chassis And Connecting Modules

    Replaceable Parts List Figure 9: Rear chassis and connecting modules MDO3000 Series Mixed Domain Oscilloscopes Service Manual...