Tektronix MDO4104-6 Instructions Manual

Mdo4000 series
Hide thumbs Also See for MDO4104-6:

Advertisement

MDO4000 Series Oscilloscopes and Application Modules
xx
Declassification and Security
ZZZ
Instructions
www.tektronix.com
*P077058401*
077-0584-01

Advertisement

Table of Contents
loading

Summary of Contents for Tektronix MDO4104-6

  • Page 1 MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions www.tektronix.com *P077058401* 077-0584-01...
  • Page 2 Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
  • Page 3: Table Of Contents

    Table of Contents Preface ......................Clear and Sanitize Procedures..................Memory Devices....................Data Export Devices................... Sanitizing an Application Module ................Built-In Security Features ..................Disable the LAN Port and Clear LAN Ethernet Settings..........Disable the USB Device Port................Clear and Sanitize a Nonfunctional Instrument ..............MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions...
  • Page 4 Table of Contents MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions...
  • Page 5: Preface

    Instrument code and calibration settings reside in nonvolatile flash memory. Instrument setups and reference waveforms may also be stored in flash memory or on USB drives connected to the instrument. If you have any questions, contact the Tektronix Technical Support Center at www.tektronix.com/support. Reference The procedures in this document are written to meet the requirements specified in:...
  • Page 6 Preface Products The following Tektronix products are covered by this document: Oscilloscopes: MDO4104-6 MDO4104-3 MDO4054-6 MDO4054-3 MDO4034-3 MDO4014-3 Application Modules: DPO4AERO DPO4AUDIO DPO4AUTO DPO4AUTOMAX DPO4COMP DPO4LMT DPO4ENET DPO4PWR DPO4USB DPO4VID MDO4TRIG MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions...
  • Page 7 Preface Related documents The Tektronix MDO4000 Series Mixed Domain Oscilloscopes Service Manual, part number 077-0585-XX, is available on the Tektronix Web site at www.tektronix.com/manuals. Terms The following terms may be used in this document: Clear. This removes data on media/memory before reusing it in a secured area.
  • Page 8 Preface MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions...
  • Page 9: Clear And Sanitize Procedures

    Clear and Sanitize Procedures The following terms are used in the tables in this section: Type of User Info Stored This column describes the type of user information that is stored in the device: User data. Waveforms and other measurement data that represent signals that users connect to the instrument.
  • Page 10: Memory Devices

    Clear and Sanitize Procedures Memory Devices The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options. Detailed procedures to clear or sanitize these devices, if any, are shown following each table. Table 1: Volatile memory devices Type user Backed...
  • Page 11 Clear and Sanitize Procedures Table 2: Nonvolatile memory devices Type user Method Type and info Data input User minimum size Function stored modification method Location accessible To clear To sanitize IC, Memory; Contains None Indirect System Main Not applicable - Not applicable - CMOS, PowerPC...
  • Page 12: Data Export Devices

    Clear and Sanitize Procedures Data Export Devices The following table lists the data export devices. Table 4: Data export devices Method Data input User Type Function modification method Location accessible To disable USB host ports Supports the Direct System USB host Files can be deleted or overwritten on the instrument or a PC;...
  • Page 13: Sanitizing An Application Module

    Clear and Sanitize Procedures Sanitizing an Application Module To sanitize an application module, follow this procedure to return the module to the default state. NOTE. This procedure is for an application module that has its license transferred to an oscilloscope. If the module license is not installed in an oscilloscope, the module is already in the default state.
  • Page 14 Clear and Sanitize Procedures MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions...
  • Page 15: Built-In Security Features

    Built-In Security Features You can use the TekSecure function to erase setup and reference waveform data stored in internal flash memory. The MDO4000 Series models have four USB host ports (two on the front and two on the rear panel) and one USB device port on the rear panel. Any USB devices can be removed and stored or destroyed.
  • Page 16 “.set” and recall the file from the Recall Setup menu. For more information on using programming commands, refer to the MDO4000 Mixed Domain Oscilloscopes Programmer Manual, Tektronix part number 077-0510-XX, available at www.tektronix.com/manuals. MDO4000 Series Oscilloscopes and Application Modules Declassification and Security Instructions...
  • Page 17: Disable The Lan Port And Clear Lan Ethernet Settings

    Built-In Security Features Disable the LAN Port and Clear LAN Ethernet Settings To disable the LAN port and clear LAN Ethernet settings, such as IP addresses, follow these steps: CAUTION. Write down all setting values before clearing them, to enable you to restore network connectivity later.
  • Page 18: Disable The Usb Device Port

    Built-In Security Features The LAN system is disabled and no longer allows data traffic in or out. The relevant LAN Ethernet settings are also cleared. Disable the USB Device Port To disable the USB device port, follow these steps: 1. Remove any USB cable or device from the USB device port on the rear of the instrument.
  • Page 19: Clear And Sanitize A Nonfunctional Instrument

    Refer to your company's internal policies regarding handling or disposal of the board. 3. Return the oscilloscope to Tektronix. A new Main board will be installed, and the oscilloscope will be repaired. The oscilloscope will be adjusted as necessary, which includes adjusting (calibrating) the new Main board and the Analog board together.