Technical Specifications - Thermo Scientific Barnstead Smart2Pure Operating Instructions Manual

Water purification system
Table of Contents

Advertisement

7. Technical specifications

Source
Blocking index (SDI)
Free chlorine concentration
Manganese content
Iron content
pH-Range
Temperature
Pressure
Conductivity µS/cm
Resistance
MΩxcm at 25°C
TOC
ppb
Bacteria
CFU/ml
Bacterial
EU/ml
endotoxines
Particles
> 0.2 µm
Flow rate
l/min
Typical amount dispensed
*
Dependent on feedwater quality and disinfection
Conductivity µS/cm
Resistance
MΩxcm at 25°C
Retention quota for bacteria
and particles
Permeate
l/h
performance
Tank
l
content
Height:
Width:
Depth:
Weight:
Smart2Pure Standard
Smart2Pure UF
Smart2Pure UV
Smart2Pure UV/UF
Sound-pressure level
12
Demands made on the feedwater
Drinking water acc. to DIN 2000.
Max. < 3 for all versions. If the value is higher, a pre-
filter must be installed upstream.
< 0.1 mg/l
< 0.05 mg/l
< 0.05 mg/l
4 - 11
2 - 35 °C
1 - 6 bar
ASTM I Product water quality (at dispensing valve)
Standard
0.055
18.2
5 - 10
< 1
--
< 1/ml
1.0
ASTM II Product water quality (in the tank)
Standard
0.067 – 0.1
10 – 15
Ø 99 %
3 or 6
approx. 6
Dimensions
Airborne sound emission
Thermo Scientific Barnstead Smart2Pure
UV
0.055
18.2
1 - 5
< 1
--
< 1/ml
1.0
5 Litres per day
UV
0.067 – 0.1
10 – 15
Ø 99 %
3 or 6
approx. 6
545 mm
305 mm
400 mm
22 kg
22 kg
22 kg
22 kg
49 db(A)
Ultrapure water system
UF
UV/UF
0.055
0.055
18.2
18.2
5 - 10
1 - 5
< 1
< 1
< 0.005*
< 0.005*
< 1/ml
< 1/ml
0.6
0.6
UF
UV/UF
0.067 – 0.1
0.067 – 0.1
10 – 15
10 – 15
Ø 99 %
Ø 99 %
3 or 6
3 or 6
approx. 6
approx. 6

Advertisement

Table of Contents
loading