Installation - Thermo Scientific Barnstead Smart2Pure Operating Instructions Manual

Water purification system
Table of Contents

Advertisement

9.2 Installation

Schwenkbares Display
Swivel display
11) Feedwater connector
2) Pure water connector
3) Concentrate connector
4) Tank water return connector Hose, 1/4" o.d.
5) Power supply connector, 4-pin, 24V DC
6) Potential-free contact connector, 5-pin
7) Connector for optional printer
8) Filter cartridge
9)
Pretreatment cartridge
10) Sterile filter
11) Sterile vent filter
Proceed as follows to install your Smart2Pure ultra pure water system and put it into op-
eration:
Stand the system at the installation location or, if it is to be wall mounted, hang it to the
wall using the wall holder that is available as accessory.
Unlock the cartridge cover by pressing the push-button and take the cover off.
Remove the stoppers from the filter cartridge and pretreatment cartridge which are
supplied and keep the stoppers for return shipment.
Fit the filter cartridge (8) in the free space at the back of the system and plug the 2
quick-connect couplings on the cartridge connectors so that they snap audibly into po-
sition.
Now fit the
pretreatment cartridge
plug the 3 quick-connect couplings on the cartridge connectors so that they snap audi-
bly into position.
Use the R 3/4" assembly kit that is supplied to make connection to the feedwater con-
nector (1).
20
Tankverschluss
Drehknopf zur Reinstwasserentnahme
Tank cap
Rotary knob for Ultra pure water
Hose, 1/4" o.d.
Hose, 1/4" o.d.
Hose, 1/4" o.d.
(9) in the free place at the front of the system and
Thermo Scientific Barnstead Smart2Pure
Ultrapure water system
Schnellverschlüsse Filterkatuschen
Quick connectors filter cartridges
11
10
8
6
7
5
3
2
9
1
4

Advertisement