Chapter 6: Troubleshooting - Respironics Tranquility Quest Plus Service Manual

Table of Contents

Advertisement

Troubleshooting Table (Continued)
Symptom
Ramp Pressure -
The pressure does not ramp
correctly.
Pressure Variation - The
pressure varies around the
set value, pressure fluctuates
greater than 1 cm H
O.
2
Compliance Meter Not
Updating
1003740
Cause
• No ramp time prescribed.
• Main PCA.
• Ramp button not
functioning.
• Internal air leak
• Filters dirty
• Main PCA, Blower
• Air path blocked
Main PCA
Tranquility Quest Plus CPAP System Service Manual
Verification
Make sure the patient's
prescription specified
ramp. Make sure unit was
set for ramp.
Replace filters
Perform Testing Process
(see Section 8.1)
Time not updating.

Chapter 6: Troubleshooting

Corrective Action
Set unit for ramp.
Replace Main PCA.
Replace Blower.
Replace Main PCA
Replace Main PCA