Chapter 6: Troubleshooting - Respironics Tranquility Quest Plus Service Manual

Table of Contents

Advertisement

Troubleshooting Table (Continued)
Symptom
Noise -
Odor -
1003740
Cause
Missing or damaged rubber
foot on bottom of unit,
blower malfunctioning.
Tubing smells new, unit
smells new, airborne residue
buildup.
Tranquility Quest Plus CPAP System Service Manual
Verification
Turn unit over and inspect
the bottom for missing or
damaged rubber feet.
Visually inspect patient
tubing for contamination.

Chapter 6: Troubleshooting

Corrective Action
Replace Blower.
Run unit in a clean environ-
ment for a few hours to
eliminate new smell. Wash
tubing with soap and water.
To clear residue build-up,
replace all subassemblies in
the patient air stream
(blower, filters, patient
circuit).
11