Tranquility Quest Plus System Final Test (Continued)
Step 7
Press the ramp button and verify that the pressure on
the manometer drops to approximately 3.0 cm H
Pressure will increase over the specified time period
until the pressure setting is reached. Verify at least 1
cm H
O in pressure change in ramp operation.
2
Record the results. If necessary, turn the unit off and
then back on again to disable the ramp feature.
Step 8
Run the unit in for four hours at the 10 cm H
setting. Record start and stop times.
Step 9
Verify the pressure to be 9.0 to 11.0 cm H
manometer after run-in. Record the results.
After run in, repeat steps 2 through 8 and then
Step 10
proceed to step 11.
Note: Perscription pressure must be reset by authorized personel.
1003740
2
O on the
2
Tranquility Quest Plus CPAP System Service Manual
0.
2
O
Chapter 8: Testing
28