Technical Data; Operating Principle - KROHNE H250H Handbook

Variable area flowmeter
Table of Contents

Advertisement

TECHNICAL DATA

8

8.1 Operating principle

The flowmeter H250 operates on the float measuring principle. The measuring unit consists of a
metal cone in which a float can move freely up and down. The medium flows through the
flowmeter from bottom to top. The float adjusts itself so that the buoyancy force B B B B , acting on it,
the form drag D D D D and its weight W W W W are in equilibrium: W = B + D.
Figure 8-1: Operating principle
1 Indication principle M9 and M8MG
2 Indication principle M10 and M8EG
For indicators M9 and M8MG 1 the flow-dependent height of the float in the measuring unit is
transmitted by means of a magnetic coupling and displayed on a scale. For indicators M10 and
M8EG 2 the flow-dependent height of the float in the measuring unit is transmitted to the
electronic display by magnetic field sensors S1 and S2.
Operating principle of H250H and H250U
Figure 8-2: Operating principle H250H and H250U
1 H250H - horizontal flow direction
2 H250U - flow direction from top to bottom
The flowmeters operate according to a modified float measuring principle. The guided float
adjusts itself so that the flow force acting on it is in equilibrium with the opposing spring force.
The flow-dependent position of the float in the measuring unit is displayed on a scale by means
of a magnetic coupling.
INFORMATION!
Flowmeters H250H and H250U only work in conjunction with indicator M9.
64
W = B + D.
W = B + D.
W = B + D.
www.krohne.com
H250
04/2009 - 4000269302 MA H250-R02

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

H250uH250/cH250/hcH250/rr

Table of Contents