7. Directions for Use
Important: Prior to every measurement, fit a new un-
damaged Probe Cover on the measuring probe. Failure
to do so may result in incorrect temperature readings.
Press the ON/OFF button. The display is activated to show
1.
all symbols for 2 seconds. (Fig. 1)
2.
The last measurement reading will be shown on the display
automatically for 3 seconds with the "M" symbol visible.
(Fig. 2)
3.
A beep sound is heard. The °C or °F symbol will flash. The
thermometer is ready to measure. (Fig. 3)
4.
Straighten the ear canal by pulling the outer ear up and back
to give a clear view of the eardrum.
Children ages 1 year to adult:
- Pull the ear up and back. (Fig. 4)
(Caution: Many pediatricians
suggest use of ear thermometers
for children older than 6 months.)
For children under 1 year:
Pull the ear straight back (Fig. 5)
5.
While pulling the outer ear, insert the probe snugly into the
ear canal and press the "START" button. Release it, when you
hear a beep sound. A beep indicates measurement is com-
plete. (Fig. 6)
6.
Remove the thermometer from the ear. The display shows the
measured temperature.
7.
Replace the probe cover after each measurement. To do this,
follow the instructions in section 6, (How to apply a new probe
cover).
8.
In order to assure accurate readings, please wait at least 30 seconds
after 3-5 continuous measurements.
-