To prevent damage to the UUT, makes sure the applied voltage to
the UUT is not larger than the rating of the UUT insulation and the
connection wires.
1. Push  to clear any output from the Calibrator.
2. Connect the UUT as in the "Connect the Calibrator to a UUT" section of this chapter.
3. Set the UUT to measure dc power on the necessary range.
4. Push the numeric keys and decimal point key to record the necessary voltage output
At voltage outputs of 100 V and above (nominal), the Calibrator usually
makes a small high-pitched sound.
5. Push  to select the polarity of the voltage (default is +).
6. Push a multiplier key, if necessary. For example, push .
7. Push .
8. The Control Display shows the amplitude of your entry. For example, 123.4567 mV
9. Push the numeric keys and decimal point key to record the necessary current output
10. Push  to select the polarity of the current (default is +).
11. Push a multiplier key, if necessary. For example, push .
12. Push .
13. The Control Display shows the amplitude of your entries. For example, 123.4567 mV
4-26
Change the voltage and current connections to this procedure.
(maximum seven numeric keys). For example, 123.4567.
(see below).
(maximum six numeric keys). For example, 234.567.
and 234.567 mA ( see below).