Fluke 5502A Operator's Manual page 209

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

x
IEEE-488
FORMAT
(Format command) Use with extreme care. Restores the contents of the nonvolatile
memory device to factory defaults. The memory holds calibration constants and setup
parameters. You lose all calibration data permanently. The CALIBRATION switch on
the rear panel of the Calibrator must be set in the ENABLE position or an execution error
occurs, except for FORMAT SETUP.
Parameter:
ALL
CAL
SETUP
Example:
FORMAT SETUP
Replace the setup parameters with the default setup values (below). (The FORMAT ALL
command is the same as FORMAT CAL and then FORMAT SETUP.) The FORMAT
SETUP command also clears the *PUD string (see the *PUD command) and SRQSTR is
set to "SRQ: %02x %02x %04x %04x" (see the SRQSTR command) and SPLSTR is set
to "SPL: %02x %02x %04x %04x" (see the SPLSTR command).
Temperature
Standard
Host Connection
GPIB Port Address
Serial Ports
EOL (end of line)
EOF (end of file)
Remote I/F
SRQSTR
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
dBm Impedance
x
IEEE-488
FUNC?
(Function query) Shows the current output, measurement, or calibration function. See the
response below for output and measurement modes.
Responses:
DCV
ACV
DCI
ACI
RES
x
x
RS-232
Sequential
(replaces the whole contents with factory defaults)
(replaces all cal constants with factory defaults)
(replaces setup parameters with factory defaults)
Features
ITS-90
GPIB (IEEE-488)
4
8 bits, 1 stop bit, xon/xoff,
parity none, 9600 baud
CRLF
012,000
term
Remote Commands
SRQ: %02x %02x %04x
%04x
Defaults
600 Ω
x
x
RS-232
Sequential
(dc volts function)
(ac volts function)
(dc current function)
(ac current function)
(ohms function)
Remote Commands
x
x
Overlapped
Coupled
Display Contrast*
Display Brightness*
RTD Power Up
Default Type
Thermocouple
Power Up Default
Type
Current Limits
Voltage Limits
*PUD string
x
x
Overlapped
Coupled
6
Commands
level 7,7
level 1,0
pt385
K
±20.5 A
±1020 V
cleared