Function List - A&D EK-300EP Instruction Manual

Ek-ep series, explosion-proof compact balance
Table of Contents

Advertisement

10-5. Function List

Class
Item
poff
func
Auto power-off
Cond
Response
st-b
Stability band
width
trc
Zero tracking
pnt
Decimal point
Cp
Comparison mode
aCai
ACAI function
Umin
Minimum unit
weight acceptable
smpl
Number of samples
ltUp
LCD Backlight
control
CpHi
Comparator upper limit
Cplo
Comparator lower limit
Unit
Weighing units to be displayed
(
)
*
Even if the weight display is "5d", the sample weight may not be accepted. This is
because the weight display data is rounded off internally.
Parameter
0
Auto power-off disabled
 1
Auto power-off enabled
0
Fast / sensitive
1
 2
3
4
Slow / stable
0
Stable when within ± 0.5d/0.5 s
 1
Stable when within ± 1d/0.5 s
2
Stable when within ± 2d/0.5 s
0
Disabled
 1
Enabled
 0
Point (.)
1
Comma (,)
 0
Comparator function disabled
1
Compares all data
2
Compares all stable data
3
Compares plus data > +5d
4
Compare stable plus data > +5d
5
Compares data > +5d or < -5d
6
Compares stable data > +5d or < -5d
0
ACAI disabled
 1
ACAI enabled
Unit weight  1d
 0
Unit weight  1/10d
1
Total sample weight  5d
2
 0
10 pcs
1
25 pcs
2
50 pcs
3
100 pcs
4
5 pcs
0
Always off
1
Turns off 3 seconds after stable
2
Turns off 10 seconds after stable
 3
Turns off 30 seconds after stable
4
Turns off 60 seconds after stable
5
Always on
Sets the upper limit value
Sets the lower limit value
Sets units to be displayed
 Factory setting
24
Description
Turns the power
OFF automatically
Good environment/
Target weighing
Priority on stability
Conditions to turn
the STABLE
indicator on
Tracks the zero
drift
Decimal point
format
Conditions to
compare.
d = minimum
display division
0: no additional
samples
d = minimum
display division
(
)
*
The number of
samples displayed
first when entered
into the unit weight
storing mode
Conditions to turn the
LCD backlight on/ off.
Weight change or
key operation
Stable = STABLE
indicator on
Refer to
"8. COMPARATOR"
Refer to "10-4. Storing
Weighing Units"

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ek-3000epEk-12kepEk ep series

Table of Contents