7-6. Counting Mode (pcs)

The balance weighs the sample pieces and calculates the unit weight. Using the
sample unit weight, the balance counts the number of objects in the sample.
 As for the minimum unit weight acceptable, refer to the function setting "Umin" of
"10-5. Function List." Please note that "Umin 1" and "Umin 2" are for counting light
objects, not for improving counting accuracy.
 The sample unit weight is stored in memory even if power to the balance is turned
OFF.
Selecting the counting mode
1. Press the MODE key to select
Storing the sample unit
2. Press the SAMPLE key to enter the sample
unit weight storing mode.
3. To select the number of samples, press the
1SAMPLE key. It may be set to 5, 10, 25, 50,
or 100.
When zero (0) is not displayed to the right side
of the number of samples, as in the figure on
the right, press the
set it to zero.
4. Place a tare container on the weighing pan, and
press the
TARE
displayed to the right side of the number of
samples.
5. Place the number of samples specified on the
pan. In this example, 25 pieces.
6. Press the ENTER key to calculate and store
the unit weight. Remove the sample.
The balance has been set to count objects with
this sample unit weight.
.
(
ZERO/ON:OFF
key. Confirm that zero is
13
:pieces)
Each pressing switches
the number of samples.
key to
Confirm the zero display
ZERO
0
10 0
25 -
Container
Weighing pan
25 0
Samples
(25 pcs)
To store
25